Influence of heat loss through probe electrical leads on thermal conductivity measurement with TPS method

 

WANG Yu-wei,  LI Yan-ning

 

National Key Laboratory of Precision Testing Techniques and Instrument, Tianjin University, Tianjin 300072, China)

 

Abstract: The transient plane source (TPS) method is developed recently to measure the thermal conductivity of materials. In the measurement, the heating power is influenced by the heat which is transferred via the probe electrical leads. This fact further influences the measurement accuracy of thermal conductivity. To solve this problem, the influence of heat loss through the electrical leads on the heating power is studied theoretically. The mathematical formula of heat loss is deduced, and the corresponding correction model is presented. A series of measurement experiments on different materials have been conducted by using the hot disk thermal constant analyzer. The results show that the influence of the heat loss on the measurement is sensitive to different test materials and probes with different sizes. When the thermal conductivity of the material is greater than 0.2 W/(m·K), the influence of the heat loss is less than 0.16%, which can be ignored. As to the lower thermal conductivity materials, it is necessary to compensate the heat loss through the electrical leads, and the accuracy of thermal conductivity measurement can be effectively improved.

 

Key words: transient plane source (TPS) method; thermal conductivity; heat loss through electrical leads; heating power

 

CLD number: TP216      Document code: A

 

Article ID: 1674-8042(2018)01-0009-07    doi: 10.3969/j.issn.1674-8042.2018.01.002

 

References

 

1] Jelle B P, Gustavsen A, Grynning S, et al. Nanotechnology and possibilities for the thermal building insulation materials of tomorrow. In: Proceedings of Zero Emission Buildings & Renewable Energy Research Conference, Trondheim, Norway, 2010: 121-132.[2] Moghaddas J, Rezaei E. Thermal conductivities of silica aerogel composite insulating material. Advanced Materials, 2016, 7(4): 296-301.

3] Jensen C, Xing C, Folsom C, et al. Design and validation of a high-temperature comparative thermal-conductivity measurement system. International Journal of Thermophysics, 2012, 33(2): 311-329.

4] Hammerschmidt U, Hameury J, Strnad R, et al. Critical review of industrial techniques for thermal-conductivity measurements of thermal insulation materials. International Journal of Thermophysics, 2015, 36(7): 1-15.

5] Zhang Y H, Zhang J T, Wang Z C, et al. Review on the thermal conductivity of insulation materials by guarded hot plate. China Metrology, 2014, (10): 82-84.

6] Chen C J, Ren D Y, Zhang L Q, et al. Measurement and analysis of thermal conductivity of rubber powder based on transient hot wire method. Polymer Materials Science and Engineering, 2015, 31(3): 118-122.

7] Coquard R, Coment E, Flasquin G, et al. Analysis of the hot-disk technique applied to low-density insulating  materials. International Journal of Thermal Sciences, 2013, 65(3): 242-253.

8] Gustafsson S E. Transient plane source techniques for thermal conductivity and thermal diffusivity measurements of solid materials. Review of Scientific Instruments, 1991, 62(3): 797-804.

9] Lin H, Du J S, Li B, et al. Thermal properties measuring of tobacco leaves based on TPS method and construction of thermal conductivity model. Journal of Henan Agricultural Sciences, 2014, 43(2): 155-160.

10] Al-Ajlan S A. Measurements of thermal properties of insulation materials by using transient plane source technique. Applied Thermal Engineering, 2006, 26(17/18): 2184-2191.

11] He Y. Rapid thermal conductivity measurement with a hot disk sensor: Part 1. Theoretical Considerations. Thermochim Acta, 2005, 436(1/2): 122-129.

12] Bohac V, Gustavsson M K, Kubicar L, et al. Parameter estimations for measurements of thermal transport properties with the hot disk thermal constants analyzer. Review of Scientific Instruments, 2000, 71(6): 2452-2455.

13] He X W, Huang L P. Verification of the measurement accuracy and application range for thermophysical properties of transient plane source (TPS) method using standard material pyroceram 9606 at room temperature. Journal of Astronautic Metrology and Measurement, 2006, 26(4): 31-42.

 

探头引线热损失对TPS法测量导热系数的影响

 

 王雨薇, 李艳宁

 

(天津大学 精密测试技术及仪器国家重点实验室, 天津 300072)

 

 :  瞬态平面热源(Transient plane source, TPS)法是一种近年来发展起来的用于测量材料导热系数的方法。 在测量过程中, 加热功率受到探头引线传热的影响, 进而会影响导热系数的测量准确度。 针对这个问题, 本文研究了测量过程中探头引线热损失对加热功率的影响, 推导了热损失的数学计算公式, 并提出了相应的修正模型。 利用hot disk热常数分析仪对不同材料进行了一系列测量实验。 实验结果表明引线热损失对测量的影响随着测量材料以及测试探头尺寸的不同而发生变化。 当材料的导热系数大于0.2 W/(m·K)时, 探头引线热损失的影响小于0.16%, 可以忽略不计; 但对于低导热系数材料的测量, 对引线热损失进行补偿可以有效地提高导热系数的测量准确度。

 

关键词:  瞬态平面热源法; 导热系数; 引线热损失; 加热功率

 

引用格式:  WANG Yu-wei, LI Yan-ning. Influence of heat loss through probe electrical leads on thermal conductivity measurement with TPS method. Journal of Measurement Science and Instrumentation, 2018, 9(1): 9-15. [doi: 10.3969/j.issn.1674-8042.2018.01.002]