此页面上的内容需要较新版本的 Adobe Flash Player。

获取 Adobe Flash Player

Design of a weak bioelectric signal acquisition circuit

ZHOU Mingjuan1, WANG Yuyuan1, RAN Li2


1. Shaanxi Railway Institute, Weinan 714000China2. Xian University of Technology, Xian 710048China


 

AbstractA surface electromyography (sEMG) signal acquisition circuit based on high-order filtering is designed. We use a two-stage adjustable amplifier and a high-order Sallen-Key bandpass filter to solve the problems of non-adjustable amplification gain and low filtering order in traditional acquisition circuits. The experimental results show that the designed sEMG signal acquisition device can eliminate power frequency interference effectively, the stopband drop of the filtering part reaches approximately -100 dB/dec, which can effectively extract useful signals between 20-500 Hz, and the amplification gain reaches 60 dB.

Key wordssurface electromyography (sEMG); two-stage amplification; high-order filtering; interference suppression; power frequency noise

 

References


1Bonfiglioli R, Botter A, Calabrese M, et al. Surface electromyography features in manual workers affected by carpal tunnel syndrome. Muscle and Nerve, 2012, 45(6)873-882

2Xu C L, Lin K, Yang C, et al. Intelligent robot collaborative control method based on calf surface electromyography. Chinese Journal of Biomedical Engineering, 2016, 35(4)385-393.

3Mi C, Chen Y, Zou L. Research on muscle fatigue detection system based on surface electromyography.Modern Electronic Technology, 2018, 41(20)78-82.

4Gao G H, Cai L, Zhang X, et al. Design and application of portable EEG signal acquisition and analysis device. Modern Electronic Technology, 2020, 43(10)21-24.

5Tang L, Ge L Q. A new method of filtering power frequency interference in surface EMG signal. Modern Electronic Technology, 2015, 38(4)30-34.

6Zhang S S, Wu W, Li T T, et al. Study on the correlation between surface electrode and monopolar needle electrode collecting tibial anterior myoelectric signal. Chinese Journal of Sports Medicine, 2019, 38(5)358-363.

7Du H L, Li X, Li S, et al. Development and preliminary clinical application of a multi-channel surface EMG signal data acquisition and analysis system for masticatory muscles. Journal of Biomedical Engineering, 2014, 31(1)23-28.

8Gao D R, Li T X, Chen Q Y, et al. High common-mode rejection ratio full-band EEG acquisition system.Chinese Journal of Biomedical Engineering, 2015, 34(6)708-713.

9Peng Z T, Xu P, Shen X Y, et al. Design of prosthetic wireless control system based on neural network.  Modern Electronic Technology, 2018, 41(2)63-67.

10Zhou B, Ji X L, Zhang R, et al. Design of surface EMG signal acquisition instrument based on embedded technology. Modern Electronic Technology, 2010, 33(6)55-57.

11Zhou M J, Lu M. Design of EMG signal acquisition circuit based on high-order filtering. Journal of Transducer Technology, 2018, 31(1)54-60.

12Zuo P F, Lu Z Y, Zhang Y Q, et al. Design of channel wireless surface EMG and acceleration signal acquisition system. Journal of Transducer Technology, 2013, 26(6)790-795.

13Mao D J, Zhang X M, Jiang X W, et al. Low-power compression filtering algorithm for surface EMG signal based on wavelet transform.Journal of Transducer Technology, 2016, 29(5)647-653.  

14Li P Y, Guo W C, Sheng X J, et al. Design of microarray surface EMG acquisition system. Journal of Sensor Technology, 2020, 33(6)916-921.

 

一种表面肌电信号采集装置的实现


周明娟1, 王语园1, 冉蠡2


1. 陕西铁路工程职业技术学院, 陕西 渭南 7140002. 西安理工大学, 陕西 西安 710048


摘要:为了更有效地采集到表面肌电信号(Surface elctromyography, sEMG), 设计了一种基于高阶滤波的sEMG信号采集装置。 两级可调放大电路及高阶Sallen-Key带通滤波器解决了传统sEMG信号采集电路中放大增益不可调、滤波阶数较低等问题。 实验结果表明, 所设计的sEMG采集装置可以很好地抑制工频干扰, 滤波部分阻带下降近似达到-100 dB/dec, 可以有效提取20-500 Hz之间的有用信号, 放大增益可达60 dB


关键词:表面肌电信号; 两级放大; 高阶滤波; 抑制干扰; 工频噪声

 

引用格式:ZHOU Mingjuan, WANG Yuyuan, RAN Li. Design of a weak bioelectric signal acquisition circuit. Journal of Measurement Science and Instrumentation, 2021, 121): 20-26. DOI103969jissn1674-8042202101003


[full text all]