Research on thermocouple transient thermometry technology based on LabVIEW


 

WANG Rui, ZHANG Zhi-jie, ZHAO Chen-yang, MU Xin-rong, ZHAI Yong

 

(Key Laboratory of Instrumentation Science &  Dynamic Measurement (North University of China),  Ministry of Education, Taiyuan 030051,  China)

 

Abstract: Based on LabVIEW platform, a distributed dynamic storage testing system is designed for measuring transient high-temperature signals of explosion field. Using a high-power semiconductor laser as heat source, a traceable dynamic calibration system is established to perform dynamic calibration of thermocouples. With quantum-behaved particle swam optimization (QPSO) algorithm on MATLAB platform, a model of dynamic compensation filter is established. It is used by LabVIEW that calls MATLAB Script nodes or COM components to accomplish the mixed programming of LabVIEW and MATLAB, further to compensate the temperature values of the thermocouples dynamically. The experimental results show that the technique that combines temperature measurement system with LabVIEW platform is applied well in testing the explosion temperature of thermobaric weaponry and makes the compensation values closer to the actual signals.

 

Key words: temperature test; dynamic calibration; MATLAB Script node; COM component; thermocouple

 

CLD number: TP274Document code: A

 

Article ID: 1674-8042(2015)02-0111-05  doi: 10.3969/j.issn.1674-8042.2015.02.002

 

References

 

[1] LI Ling. Principle and implementation of thermobaric bomb. Modern Military, 2005, (3): 55-57.
[2] WANG Dai-hua, SONG Lin-li, ZHANG Zhi-jie. Contact measuring method of explosion temperature based on tungsten-rhenium thermocouple. Journal of Detection & Control, 2012, 34(3): 23-28.
[3] YANG Lei, FANG Jian-cheng, ZHAO Zi-yu, et al. Transient temperature measurement system based on LabVIEW. Instrument Technique and Sensor, 2008, (6): 49-51.
[4] HAO Xiao-jian. Transient high surface temperature measurement and dynamic calibration. Taiyuan: North University of China, 2005.
[5] MENG Xie-wen. Research on dynamic measurement errors compensation technology of transient surface temperature sensor. Taiyuan:  North University of China, 2007.
[6] Arwatz G, Bahri C, Smits A J, et al. Dynamic calibration and modeling of a cold wire for temperature measurement. Measurement Science and Technology, 2013,  24(12):  1-11.
[7] LI Fei, ZHENG Heng. Application of composite programming of LabVIEW and MATLAB to test of digital antenna array. Measurement & Control Technology, 2007, 33(3): 1-3.
[8] JING Su-tong, ZHAO Fu-tang. Thermocouple temperature measure system based on LabVIEW. Instrument and Metrological Technology, 2005, (6): 37-39.
[9] WU Jian, ZHANG Zhi-jie. Correction of sensor’s dynamic error caused by system limitations. Journal of Measurement Science and Instrumentation, 2012, 3(1):  75-79.
[10] SONG Guang-dong, WANG Chang, WANG Jin-yu,  et al. Implemening mixed programming with LabVIEW and MATLAB based on DLL technology and COM component technology. Computer Application and Software, 2013, 30(1):  287-289.
[11] YAO Juan. The design of software platform of distributed measurement system based on WALN. Taiyuan: North University of China, 2013.
[12] HUANG Shi-kui, PENG Yu-ning, XIE Shu-ping, et al. Methods and application of mixed programming with LabVIEW and Matlab. Research and Exploration in Laboratory, 2009, 28(7):  67-71.

 

基于LabVIEW的热电偶瞬态测温技术研究

 

王瑞, 张志杰, 赵晨阳, 穆欣荣, 翟永

 

(中北大学 仪器科学与动态测试教育部重点实验室, 山西 太原 030051)

 

摘要:采用分布式动态存储测试技术, 构建了基于LabVIEW平台的爆炸瞬态高温测试系统。 选用大功率半导体激光器为热源的可溯源动态校准系统对热电偶进行动态校准, 应用量子粒子群优化算法设计了动态补偿滤波器模型, 并通过LabVIEW调用MATLAB Script节点或COM组件完成LabVIEW与MATLAB的混合编程, 实现了对热电偶测温系统输出信号的动态补偿, 使之更接近真实值。 实验结果表明, 该测试系统与LabVIEW平台结合的温度测量技术很好地适用于对温压类武器爆炸过程的瞬态温度测试。

 

关键词:温度测试; 动态校准; MATLAB Script节点; COM组件; 热电偶

 

引用格式: WANG Rui, ZHANG Zhi-jie, ZHAO Chen-yang, et al. Research on thermocouple transient thermometry technology based on LabVIEW. Journal of Measurement Science and Instrumentation, 2015, 6(2):  111-115. [doi: 10.3969/j.issn.1674-8042.2015.02.002]

 