此页面上的内容需要较新版本的 Adobe Flash Player。

获取 Adobe Flash Player

Effects of pinholes on zerooffset of single-axis normal-tracking measurement system

 

LI Huai-lu1, WANG Zhong1, KANG Yan-hui2, FU Lu-hua1
 
 
(1. State Key Laboratory of Precision Measuring Technology and Instrument, Tianjin University, Tianjin 300072, China;2. Division of Metrology in Length and Precision Engineering, National Institute of Metrology, Beijing 100029, China)
 
 
Abstract: A single-axis normal-tracking measurement system is proposed, which can solve the problem of measuring large curved surface. According to Collins formula, the tilt dependent error of the measurement system is studied, which uses Gaussian beam as the light source. By theoretical analysis and numerical simulation, the influence of the error is presented. The results show that there is the difference between point source and Gaussian beam for differential confocal microscopy. The optimal diameter of pinhole can be determined by the mathematical model and the actual parameters of the measurement system. The optimal pinhole diameter of this measurement system is 20 to 35 for 633 nm wavelength light source.
 
 
Key words: differential confocal microscopy; Gaussian beam; Collins formula; pinhole optimum
 
 
CLD number: TP391.4Document code: A
 
 
Article ID: 1674-8042(2016)04-0336-06doi: 10.3969/j.issn.1674-8042-2016-04-005
 
 
 
References
 
 
[1]HE Yan-wen, CHENG Yi, TANG Zhi-lie. Imaging investigation of polarization control microscope based on stokes parameters of weak anisotropic substances. Acta Laser Biology Sinica, 2016, 44(1): 21-26.
[2]ZHU Zu-song, ZHANG Jie, YI Ming-fang, et al.Preparation of the high-quality and stable microcrystalline silicon films. Journal of Sichuan University(Natural Science Edition), 2016, 53(1): 157-162.
[3]MA Yu-fan, SHENG Lei-mei. Study on the preparation and performance of multi-walled carbon nanotubes as conductive agent of lithium ion battery.New Chemical Materials, 2016, 44(1): 43-45.
[4]WEI Tong-da, ZHANG Yun-hai, XIAO Yun, et al. Effects of polarization state and effective numerical aperture on the resolution in confocal total internal reflection microscopy. Laser & Optoelectronics Progress, 2014, 51(1): 315-321.
[5]Qiu L, Liu D, Zhao W, et al. Real-time laser differential confocal microscopy without sample reflectivity effects. Optics express, 2014, 22(18): 21626-21640.
[6]ZHANG Jian-huan, TAN Jiu-bin. Theoretical analysis of surface slope measurement with a confocal microscope.Journal of Optoelectronics Laser, 2002, 13(12): 1272-1275.
[7]HOU Mao-sheng. Study on key techniques of profile measurement for large aperture optical aspherics, 2015.
[8]GAO Yu-han. Research of Gaussian beam shaping. Changchun:Changchun University of Science and Technology, 2012.
[9]GE Hua, YANG Kun-tao. The research of conversion of off-axis Gaussian beam passing through optical system. Laser & Infrared, 2003, 33(1): 68-70.
[10]ZHU Lei-ming, WU Xiao-ping, LI Jian-wei, et al. The eulers rotation and dynamic equation of rectangular coordinate system. Hydrographic Surveying and Charting, 2010, 30(3): 9-12.
[11]DING Gui-lin. Collins formula and its applications in optics. Chinese Journal Of Quantum Electronics, 1990, 7(3): 221-227.
 
 
针孔对单轴法线跟踪式测量系统零位测量的影响
 
 
李怀禄1, 王仲1, 康岩辉2, 付鲁华1
 
 
(1. 天津大学 精密测试技术与仪器国家重点实验室, 天津 300072; 2. 中国计量科学研究院 长度计量科学与精密机械测量技术研究所,北京 100029)
 
 
摘要:针对大曲率表面面形测量问题, 提出一种单轴法线跟踪式测量方法。 通过运用Collins公式, 研究了倾斜角度对基于高斯光束测量系统的影响;通过理论分析和数值模拟, 证明了在待测面倾斜的情况下, 高斯光束与点光源作用下差动共焦系统的零位位置有明显不同; 推导出高斯光束测量系统中针孔最佳尺寸和实际系统参数的关系,指出实验中波长为633 nm测量系统的针孔最佳直径在20 到35 之间。
 
 
关键词:差动共焦; 高斯光束; Collins公式; 针孔最佳尺寸
 
 
引用格式:LI Huai-lu, WANG Zhong, KANG Yan-hui, et al. Effects of pinholes on zerooffset of single-axis normal-tracking measurement system. Journal of Measurement Science and Instrumentation, 2016, 7(4): 336-341. [doi: 10.3969/j.issn.1674-8042.2016-04-005]